Chapter 7
Ventilator Communications
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
7-5
To program the serial number only (without downloading software):
1. Enter Esprit diagnostic mode: simultaneously press the
ALARM
RESET
and
100% O
2
keys on the front panel for approximately five
seconds while turning ventilator power on. A warning to verify that no
patient is connected. Touch
OK
to enter diagnostic mode.
2. With the Esprit software CD-ROM in the CD drive, click on
Start >
Run
, then enter
D:\setup.exe -snonly
to begin a software download,
where
D:
is the CD drive letter. Enter the command exactly as shown,
in lowercase letters with a space between
exe
and
-snonly
.
3. Click
Next
at each of the dialog boxes until you are prompted to enter
the serial number.
4. When a dialog box prompts you to do so, enter the ventilator serial
number (
VSxxxxxxx
)
after
the ventilator has restarted in diagnostic
mode, where
xxxxxxx
is the seven-digit ventilator serial number. Enter
VS
in uppercase letters.
Enabling Options
CAUTION:
Option enable buttons can be used
one time only
: do
not
install options
until you have verified that the ventilator serial number is displayed at
power up or on the software screen in diagnostic mode. If the serial
number must be reprogrammed, be sure to do so
before
enabling an
option.
To enable an option, a parallel port adapter (P/N 1004644) and option enable
buttons are required. Follow these steps:
1. Insert the option button (label side up) into the parallel port adapter
(Figure 7-5), then plug the adapter into the parallel port on the
ventilator back panel.
2. Power up the ventilator in diagnostic mode: the ventilator
automatically reads the option from the option enable button.
3. To confirm that the option is enabled, verify that the option appears in
the software screen.