Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-95
14. Disconnect the original color display cable from the DC/DC converter
PCBA, the intensity control potentiometer, the main PCBA, and the
ground lug from the shielding plate. Remove and discard the original
color display cable.
15. Remove the DC/DC converter PCBA from the power supply shroud and
discard.
16. Route the color display data cable and the backlight inverter cable
through the shielding plate into the top enclosure. Reconnect the
remaining ribbon cables to the GUI.
17. Carefully install the GUI onto the top enclosure. Insert one M3 x 8-mm
screw into one M3 lock washer. Hand tighten the screws.
18. Install the adapter PCBA onto the main PCBA. Verify the adapter
PCBA is fully seated. Insert one screw with lock washer through the
adapter PCBA. Tighten to 4.0 ± 1 in-lbs. (Figure 9-101).
Figure 9-101: Installing the Adapter PCBA
19. Connect the backlight inverter cable to the intensity control
potentiometer. Verify the cable is fully seated.
20. Install the color display data cable and the backlight inverter cable
onto the adapter PCBA. Verify the cables are fully seated.
Installing the CPU PCB
Remove the existing CPU PCB and replace with the CPU PCB included in the
upgrade kit (see “Printed Circuit Boards (Daughter PCBs) (Except Main PCB)”
on page 9-44).