Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-91
5. Reset the altitude and compliance compensation settings in the user
configuration screen.
6. Perform an EST with the top enclosure removed. If the EST passes,
install the top enclosure (see “Top Enclosure” on page 9-4).
Enabling Options
Enable an option (see Chapter 7) only if the option upgrade is being installed
at this time.
Final Checkout
Complete the recommended tests in the performance verification procedure
(see Chapter 8).
Replacing the GUI
(Original 10.4-in. to
2nd Generation
10.4-in)
1. Disconnect power, unscrew and separate the ventilator top and bottom
halves. Set the halves side by side.
2. Remove the GUI assembly from the enclosure (see “GUI Assembly
(Original 10.4-in. GUI)” on page 9-21).
3. Remove the four screws and washers securing LCD to the LCD bracket
Figure 9-95: Disassembling the LCD