Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-11
7. Lift the power supply far enough to disconnect the cables from
connectors J1, J2, and J3, then lift the power supply from the top
enclosure.
Figure 9-14: Power Supply Wire Connections
TB1 pins: 1, 2, 3 and 4
TB2 pins: 1, 2, 3 and 4
Do not remove ground wire
from pin 3.
TB3 pins
(Do not remove ground wire from pin 3.)
4
3
2
1
Connector
TB3
Connector
TB1
Connector TB2
(underneath PCB)