Chapter 9
9-26
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-28: Volume and Intensity Potentiometers (9.5-in. GUI)
Intensity and Volume
Potentiometers
(10.4-in. GUI’s)
Follow these steps to remove the intensity and volume potentiometers (Figure
9-29). Reverse removal steps to install as shown.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Unscrew the two 2.5-mm screws from the bottom corners of the GUI.
3. Remove the GUI knob cover, loosen the 11-mm nut, then remove the
knob.
4. Remove the ½-in. nut and star washer to the rotary encoder.
5. Separate the GUI from the enclosure. Disconnect all interconnect
harnesses between the GUI and enclosure.
6. Loosen the set screws (one for each knob) holding the brightness and
alarm volume knobs.
7. Remove the ½-in. nut and washer holding each potentiometer, then
disconnect the harnesses and remove the potentiometers.
Volume potentiometer (connected to harness VC/POT)
Intensity potentiometer (connected to harness IC/POT
Set screws