Chapter 8
Performance Verification
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
8-57
Analyzer air reads 900 - 1100 mL (square
wave)
mL
mL
mL
Analyzer air reads 900 - 1100 mL (ramp
wave)
mL
mL
mL
Analyzer air reads 2250 - 2750 mL (ramp
wave)
mL
mL
mL
Analyzer air reads 2250 - 2750 mL
(square wave)
mL
mL
mL
Analyzer O
2
reads 2250 - 2750 mL
(square wave)
mL
mL
mL
Analyzer O
2
reads 2250 - 2750 mL (ramp
wave)
mL
mL
mL
Analyzer O
2
reads 900 - 1100 mL (ramp
wave)
mL
mL
mL
Analyzer O
2
reads 900 - 1100 mL (square
wave)
mL
mL
mL
Analyzer O
2
reads 450 - 550 mL (square
wave)
mL
mL
mL
Analyzer O
2
reads 450 - 550 mL (ramp
wave)
mL
mL
mL
Analyzer O
2
reads 225 - 275 mL (ramp
wave)
mL
mL
mL
Analyzer O
2
reads 225 - 275 mL (square
wave)
mL
mL
mL
Test 10: Oxygen Accuracy
Passed
Value
Failed Value
Corrected
(after repair)
Value
External O
2
monitor reads 18 - 24% (VCV
mode)
%
%
%
Ventilator O
2
display reads 18 - 24% (VCV
mode)
%
%
%
External O
2
monitor reads 27 - 33% (VCV
mode)
%
%
%
Ventilator O
2
display reads 27 - 33% (VCV
mode)
%
%
%
External O
2
monitor reads 57 - 63% (VCV
mode)
%
%
%
Ventilator O
2
display reads 57 - 63% (VCV
mode)
%
%
%
External O
2
monitor reads 77 - 83% (VCV
mode)
%
%
%
Ventilator O
2
display reads 77 - 83% (VCV
mode)
%
%
%
External O
2
monitor reads 97 - 103% (VCV
mode)
%
%
%
Test 9: Gas Volume Accuracy
Passed Value
Failed Value
Corrected
(after repair)
Value