Chapter 6
6-22
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Diagnostic Codes
7xxx: Power Supply Failures (Vent Inop. - All)
NOTE:
A 7xxx code that precedes a 1008 or 1014 in the diagnostic log
indicates that the power supply output voltage has dropped below
tolerance three times. Always confirm fuse integrity before replacing
power supply.
7000
Power failure, power fail.
Replace power supply.
7002
Power failure, power fail.
1. If code 7002 occurs in conjunction
with code 1008 or 1014, slave in
replacement power switch and
replace if the problem is resolved.
2. Replace power supply.
7004-7014
Various combinations of +12V, -12V,
and power fail failures.
Replace power supply.
7016-7030
Various combinations of 24V, +12V,
-12V, and power fail failures.
1. Confirm power cord is fully
connected.
2. Replace power supply.
7032-7062
Sensor PCB +5V failure and/or
various combinations of +12V, -12V,
24V, and power fail failures.
1. Replace sensor PCB.
2. Replace power supply.
7064-7094
MMI PCB +5V failure and/or various
combinations of +12V, -12V, 24V,
and power failures.
1. Replace MMI PCB.
2. Replace power supply.
7096-7126
MMI PCB +5V failure and/or various
combinations of sensor PCB +5V,
+12V, -12V, 24V, and power fail
failures.
1. Replace MMI PCB.
2. Replace sensor PCB.
3. Replace power supply.
7128-7158
Various combinations of +10V,
+12V, -12V, 24V, and power fail
failures.
Replace power supply.
7160-7190
+10V failure and various
combinations of sensor PCB +5V,
+12V, -12V, 24V, and power fail
failures.
1. Verify power supply harness
connections.
2. Replace sensor PCB.
3. Replace power supply.
7192-7222
+10V failure and various
combinations of MMI PCB +5V,
+12V, -12V, 24V, and power fail
failures.
1. Verify power supply harness
connections.
2. Replace MMI PCB.
3. Replace power supply.
7224-7254
+10V failure and various
combinations of MMI PCB +5V,
sensor PCB +5V, +12V, -12V, 24V,
and power fail failures.
1. Verify power supply harness
connections.
2. Replace sensor PCB.
3. Replace MMI PCB.
4. Replace power supply.
8xxx: Software Faults
Diagnostic
Code
Code Description
Recommended Repair