Chapter 9
9-86
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
11. Attach the new shielding plate and conductive tape (Figure 9-92).
12. Feed the replacement harness through the enlarged bulkhead hole and
connect the ground wire to the threaded stud using the supplied 7-
mm nut.
NOTE:
NOTE: Each of the three ball catches that hold the GUI to the
enclosure has two tension adjustment screws. Adjust the ball
catch tension by first tightening the tension screws fully
clockwise, then unscrewing counterclockwise one and a quarter
turn.
Figure 9-92: Front of Ventilator Before Installing 10.4-in. GUI
13. Position the foil blanket in the enclosure and reattach the bulkhead
plate and MMI PCB.
14. Connect all of the interconnect cables to the 10.4-in. GUI and
reinstall the potentiometer knobs with the set screws toward the face
of the GUI (on 9.4-in. GUIs the knobs were installed with set screws
toward the inside of the enclosure). Ensure that the knobs don’t bind
or rub when turned completely in either direction.
15. Align the foil shield holes with the power supply mounting holes.
Reinstall the power supply. Reconnect J2 and J3 harness connections.
Ball catches
Reinstall
conductive tape
Shielding plate
Reinstall knobs
with set screws
facing away from
enclosure
Conductive tape
installed under
cable
Reinstall
conductive tape
Ball catch