Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-69
Oxygen Regulator
Assembly (with
Oxygen Pressure
Switch)
Follow these steps to remove the oxygen regulator assembly (Figure 9-80):
1. Remove top enclosure (see “Top Enclosure” on page 9-4).
2. Remove enclosure brace.
3. Remove oxygen flow sensor (see “Oxygen Flow Sensor” on page 9-57),
but do not cut the tie wrap or disconnect the O2 F/S cable from the
flow sensor.
4. Remove the air flow sensor but do not cut the tie wrap or disconnect
the AIR F/S cable from the flow sensor (see “Air Flow Sensor” on
page 9-65).
5. Remove the air valve assembly (see “Air Valve Assembly” on page 9-
66).
6. Remove the oxygen valve assembly (see “Oxygen Valve Assembly” on
7. Use a 4-mm hex wrench to remove the two screws holding the oxygen
water trap to the bracket.
8. Use a 2.5-mm hex wrench to remove the bracket screw above the
water trap.
9. Unscrew the bowl from the oxygen water trap, then unscrew the water
trap from the inlet block.
10. Remove the four 2.5-mm screws holding the oxygen regulator
assembly to the back plate.
11. From inside the enclosure, disconnect the two black wires from
connectors J5 and J6 on the sensor PCB.
12. Carefully cut the tie wraps holding the two black wires near the sensor
PCB, then carefully separate these wires all the way back to pressure
switch 1.
13. Lift the oxygen regulator assembly away from the enclosure.