Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Diagnostic Codes
pressure when deenergized, without displaying any pressures between
0 cmH
2
O and PRV cracking pressure.
7. If the difference between the
Inhalation Pressure
and
Exhalation
Pressure
displays on the Hardware screen is greater than 2 cmH
2
O at
any time, note which is higher, then swap the INSP and EXH tubes on
the sensor PCB and retest.
•
If the pressures are now higher on the opposite pressure display,
replace the 3-station solenoid assembly.
•
If pressures remain higher on the same pressure display, replace
the sensor PCB.
8. Return the INSP and EXH tubes to their correct positions on the
sensor PCB.