The blower motor has seized or is drawing too much current.
•
The blower controller PCB is not operating properly.
•
The air flow sensor is not reading accurately.
•
The power supply is not providing the 29 VDC (24 VDC if running on
battery) output to the blower controller PCB.
•
The sensor PCB is not operating properly.
•
Faulty harness connections between the blower controller PCB and
power supply or sensor PCB.
Follow these steps to determine the cause and repair for a 1012 diagnostic
code:
1. Determine if the blower is working: listen for the blower to start
spinning at startup, or use the Hardware screen (section 5) to set an
air flow: if there is flow, the blower is working.
2. If the blower is not working, determine if the cooling coil fan on the
bottom of the unit is operating by looking, feeling for air flow, or
carefully probing with a slender flexible object such as a wire or tie
wrap. (The blower and this cooling fan share the same output from the
power supply.)
3. If both the blower and the fan are not working, check for 29 VDC (24
VDC if running on battery) at the blower controller PCB:
a.
Enter diagnostic mode.
b.
Place DMM leads on the red and black wires on the green terminal
block on the blower controller PCB mounted next to the blower.
c.
Make sure the wires to the terminal block are securely connected.
Check the other end of this harness and verify it is securely
connected to power supply connector J1.
d.
If F1 and/or F2 are open, replace fuses.
e.
If 24/29 VDC is not present, replace the power supply.
4. If the blower is not working but the fan is, check the LED (which
indicates an overcurrent condition) on the blower controller PCB
during a normal startup attempt. If the LED lights, either the blower
motor or the blower controller PCB is not functioning correctly.
a.
If you can hear the blower attempt to start running, and then the
LED lights, replace the blower.