Appendix A
A-12
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Crossover Circuit
(Test 5, Step 2)
This step pressurizes the circuit with oxygen to verify that the crossover
solenoid (SOL1) operates properly when de-energized.
Diagnostic code
3117
indicates a probable SOL1 failure.
Red = air flow
Blue = O
2
flow
Black = no flow
Содержание ESPRIT