Appendix A
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
A-15
Oxygen Sensor
(Test 7, Step 1)
This step establishes 100% oxygen at 200 LPM for one second, and 1 LPM for
two minutes, to calibrate the oxygen sensor.
Diagnostic code
3113
indicates a bad oxygen sensor or oxygen sensor cable.
The ventilator must pass EST following oxygen sensor replacement.
Red = air flow
Blue = O
2
flow
Black = no flow