Chapter 8
8-46
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
Test 11: Heated Exhalation Bacteria Filter, Power Failure Alarm,
and Display Intensity
Test 12: Neonatal Option Test
Symptom
Recommended Repair
Backup audible alarm
fails to sound
1. Slave in replacement backup alarm.
2. Replace the MMI to backup alarm cable.
3. Replace digital PCB.
4. Replace MMI PCB.
Backup audible alarm
fails to sound for at
least two minutes
1. Replace main PCB.
2. Replace digital PCB.
Filter heater is not warm
to the touch after 20
minutes of operation
1. Verify that the 26-29 VDC is present on heater connector
terminals (heater connector is on the main to motor 1, 2, 3 cable).
2. Replace filter heater assembly.
3. Replace the main PCB.
Power switch turned off
but backup alarm
remains on
1. Slave in replacement power switch.
2. Replace power switch harness.
3. Replace digital PCB.
Display not visible when
set to minimum
intensity
1. Replace backlight inverter harness.
2. Replace backlight inverter PCB.
3. Replace LCD.
Symptom
Recommended Repair
Inhalation/exhalation
pressure readings out of
range at 5 cmH
2
O
1. Check for leaks at circuit connections, test lung, filters, etc.
2. Check for kinked or cut tubing from inhalation module to SOL4,
and from SOL4 to sensor PCB.
3. Check for kinked or cut tubing from exhalation pressure tap to
SOL3, and from SOL 3 to sensor PCB.
4. Check for leaks at the oxygen sensor/oxygen valve connection.
5. Replace the sensor PCB.
6. Replace the 3-station solenoid.
Low flow rate out of
range
1. Confirm correct altitude setting.
2. Check that safety valve solenoid is energized.
3. Verify that CV3 is correctly positioned and oriented.
4. Check for exhalation bacteria filter occlusions (perform the back
pressure test or replace filter).
5. Verify that CV4 is correctly positioned and oriented. If not,
reposition and re-test.
6. If CV4 is correctly positioned and oriented, replace CV4.
7. Replace the exhalation flow sensor.