Chapter 8
8-16
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
9. Check the
Altitude
setting and correct it if necessary.
Write down the
altitude:
the oxygen flow and pressure accuracy tests require correct
altitude information.
10. EST is complete.
Air Flow Accuracy (Test 3)
The air flow accuracy test verifies the accuracy of the air and exhalation flow
sensors and function of the air valve.
1. If the ventilator is not already in diagnostic mode, simultaneously
press the
ALARM RESET
and
100% O
2
keys on the front panel for
approximately five seconds while turning ventilator power on, then
touch
OK
to enter diagnostic mode.
2. Disconnect the oxygen source from the ventilator.
3. Touch
Safety
on the diagnostic mode Hardware screen to energize
(white background) the safety solenoid (Figure 8-15).
Figure 8-15: Diagnostic Mode - Hardware Screen
Air
position
Safety
solenoid
energized