Chapter 8
8-18
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
9. The air flow accuracy test is complete.
Figure 8-17: Flow Accuracy Test Configuration
Oxygen Flow Accuracy (Test 4)
The oxygen flow accuracy test verifies the accuracy of the oxygen flow sensor
and function of the oxygen valve.
1. If the ventilator is not already in diagnostic mode, simultaneously
press the
ALARM RESET
and
100% O
2
keys on the front panel for
approximately five seconds while turning ventilator power on, then
touch
OK
to enter diagnostic mode.
2. Reconnect the oxygen source to the ventilator.
165 LPM
148.5 to 181.5 LPM
0 LPM
0 ± 0.1 LPM
Touch
Air
and set the
flow to:
Check that the analyzer and
Air Flow
and
Exhalation Flow
displays read:
Содержание ESPRIT