Chapter 5
5-22
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
a.
Turn the pressure relief valve hex cap clockwise to decrease or
counterclockwise to increase, so that analyzer reads 130-140
cmH
2
O.
b.
Touch
Safety
to relieve system pressure, then Touch
Safety
to re-
pressurize.
c.
Check that the analyzer reads 130-140 cmH
2
O. Repeat steps
a
to
c
if necessary.
7. Return the hardware screen to its initialized state.
Figure 5-14: Pressure Relief Valve Hex Cap
Crossover Solenoid
1. Disconnect the oxygen source from ventilator.
2. Touch
Oxygen
and set the flow to 100 LPM, then back to 0 LPM to
relieve any trapped oxygen pressure.
3. Connect patient circuit to ventilator and block wye with a cork.
4. Touch
Safety
to energize (white background) the safety solenoid.
5. Touch
Exhalation
and set it to 2000 steps.
Hex cap