Chapter 8
8-58
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
NOTE:
Neonatal Option testing is not performed if neonatal option is not
installed.
Ventilator O
2
display reads 97 - 103% (VCV
mode)
%
%
%
External O
2
monitor reads 93 - 99% (PCV
mode)
%
%
%
Ventilator O
2
display reads 93 - 99% (PCV
mode)
%
%
%
External O
2
monitor reads 57 - 63% (PCV
mode)
%
%
%
Ventilator O
2
display reads 57 - 63% (PCV
mode)
%
%
%
External O
2
monitor reads 22 - 28% (PCV
mode)
%
%
%
Ventilator O
2
display reads 22 - 28% (PCV
mode)
%
%
%
Test 11: Heated Exhalation Filter, Power Fail Alarm, and Display
Intensity
Circle One
Display information still visible with intensity knob set to minimum
PASS
FAIL
Audible alarm sounds
PASS
FAIL
Exhalation bacteria filter is warm to the touch
PASS
FAIL
Audible alarm sounds after two minutes
PASS
FAIL
Audible alarm was silenced
PASS
FAIL
Test 12: Neonatal Option Testing
Test 12a: Extended Self Test (EST) in Neonatal Mode
Circle One
Did EST pass in Neonatal mode?
PASS
FAIL
Test 12b: Pressure Accuracy
Initial
Value
Passed
Value
Failed
Value
Corrected
(after
repair)
Analyzer pressure display 4-6 cmH
2
O
cmH
2
O
Inhalation Pressure display = analyzer
display ± 10%
cmH
2
O
cmH
2
O
cmH
2
O
Exhalation Pressure display = analyzer
display ± 10%
cmH
2
O
cmH
2
O
cmH
2
O
Test 10: Oxygen Accuracy
Passed
Value
Failed Value
Corrected
(after repair)
Value