Chapter 8
8-34
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
Heated Exhalation Bacteria Filter, Power Fail Alarm, and
Display Intensity (Test 11)
This test verifies the integrity of the exhalation filter heater system, the power
fail alarm, and the display intensity.
WARNING:
Use caution when removing the filter: the heater conductor may be hot.
Start here if the optional External Battery is installed:
1. Turn ventilator power off.
2. Disconnect the External Battery cable from the back panel of the
ventilator.
3. Turn the ventilator on.
4. Verify display information is still visible when intensity knob is set at
minimum.
5. With the ventilator turned on, unplug the power cord from the AC
outlet.
6. Check that the audible alarm sounds and start the stopwatch.
7. Remove the exhalation bacteria filter from the ventilator.
8. Check that it is warm to the touch.
9. Reinstall the exhalation bacteria filter.
10. Check that the audible alarm is still sounding after 2 minutes.
11. Turn ventilator power off.
12. Check that the audible alarm is silenced.
13. Plug the ventilator into an AC outlet.
14. The heated exhalation bacteria filter and power fail alarm test is
complete.
Start here if only the Backup Battery is installed:
1. Turn ventilator power off.
2. Disconnect the Backup Battery cable from the back panel of the
ventilator.
3. Turn the ventilator on.