Chapter 5
5-14
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
voltage set to 4 V (blower voltage may vary depending on the altitude
setting), Blower on, Filter Heat off, 24V Power on, Crossover solenoid
energized (white background), and all other solenoids de-energized (gray
background).
NOTE:
At the end of each test, return the hardware screen to its initialized state.
NOTE:
The pneumatic calibration analyzer should always be in BTPS when
measuring volume, and in ATP when measuring flow.
NOTE:
Use a known-good reusable patient tubing circuit and reliable exhalation
bacteria filter.
Oxygen Valve
1. Connect the oxygen source to the ventilator.
2. Disconnect the bacteria filter, patient circuit tube, and oxygen sensor
tee from the ventilator inspiratory outlet.
3. Touch
User Config
and note the altitude.
4. Touch
Hardware
.
5. Touch
Safety
to energize (white background) the safety solenoid.
NOTE: The sensor readings on an actual Hardware screen vary from
those shown in Figure 5-10.
6. Touch
Oxygen
and set the flow to 1 LPM.
7. Check that the
Oxygen Position
display is between 185 and 525
steps.
8. Touch
Oxygen
and set the flow to 180 LPM.
a.
Verify that the
O2 Position
display corresponds to the set altitude:
Ventilator Altitude
Setting
Oxygen Position
Display
0 - 2000 ft.
1175 - 1575 steps
2001 - 4000 ft.
1145 - 1540 steps
4001 - 6000 ft.
1120 - 1505 steps
6001 - 8000 ft.
1090 - 1470 steps
8001 - 10,000 ft.
1060 - 1435 steps