Appendix A
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
A-19
Exhalation Valve
(Test 10, Step 2)
This step sets various circuit pressure levels to verify that the exhalation valve
can control PEEP accurately.
Diagnostic code 3108 indicates that exhalation pressure is out of range.
This
can be due to incorrect altitude setting, damaged exhalation filter,
nonstandard circuit size, or faulty exhalation valve/stepper motor controller
PCB.
Red = air flow
Blue = O
2
flow
Black = no flow