Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-51
3. Disconnect cable SEN J7 from the sensor PCB.
4. Remove the 8-mm nut and washer that hold the FIO
2
connector, then
remove the connector.
Figure 9-61: FIO
2
Connector
Exhalation Flow
Sensor
Follow these steps to remove the exhalation flow sensor (Figure 9-62, Figure 9-
63):
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the enclosure brace.
3. Cut the tie wrap holding the exhalation flow sensor to the exhalation
valve, and the tie wrap holding cable EXH F/S to the exhalation flow
sensor.
4. Disconnect cable EXH F/S from the exhalation flow sensor.
5. Remove the screw holding the exhalation outlet port to the exhalation
valve.
6. Remove the screw holding the outlet port to the bottom enclosure.
7. Push the flow sensor toward the vent port.
8-mm nut
and washer