Chapter 8
8-60
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
Backup Battery fully charged?
PASS
FAIL
N/A
Backup Battery is
not
fully charged if Battery Low indicator lights during normal operation on AC
power. Performance verification result is
LIMITED USE
Did the Performance Verification Pass?
(Circle One)
YES
NO
LIMITED USE (Backup Battery)*
LIMITED USE (External Battery)*
* LIMITED USE indicates that the Backup Battery or External Battery operational check failed.
Recommend that the operator connect the Ventilator to AC power for at least 12 hours and verify
battery performance.
Technician’s Signature
Date
Test 13: Backup Battery and External Battery Test
Circle One