Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-15
Figure 9-19: Pinch Fuse Clips Together
MMI PCB
(9.5-in. GUI)
Follow these steps to remove the MMI PCB (Figure 9-20) for ventilators with
9.5-in. GUI. Reverse removal steps to install.
NOTE:
The GUI assembly must be installed before reinstalling the MMI PCB.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the power supply shroud (see “Power Supply Fan/Shroud” on
3. Remove the screws securing the power supply to the top enclosure.
4. Move the power supply out of the way to allow access to the MMI
mounting screws.
5. Remove the sensor PCB (see “Sensor PCB” on page 9-5).
6. Remove the cables from the connectors on the MMI PCB:
Remove cables in this order:
From this MMI PCB
connector:
Ribbon cable from VGA display*
Reinstall with black side facing MMI PCB
J7
MMI J4
PRIMARY ALARM
J4
* When removing, lift connector lock before pulling cable. Do not use a tool.