Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-29
Figure 9-31: Rotary Encoder (10.4-in. GUI)
GUI Front Panel
Overlay (10.4-in.
GUI’s)
Follow these steps to replace the front panel overlay assembly (Figure 9-32,
Figure 9-33):
1. Remove the GUI assembly from the enclosure (see “GUI Assembly
(Original 10.4-in. GUI)” on page 9-21).
2. Remove the language inserts from the front panel. Use a small flat
screwdriver to lift the inserts, then pull to remove.
3. Use a small flat head screwdriver to lift the corner of the keyboard
assembly, then peel the keyboard from the bezel.
4. Use a soft cloth and isopropyl alcohol to remove any adhesive
remaining on the bezel.
5. Peel the backing off the replacement overlay, and remove the small
pieces of back behind the ribbon cables.
6. Thread the ribbon cables through the bezel slots, then align the
overlay over the bezel and adhere with overlay centered in the bezel
frame.
7. Reinstall language inserts. Use a pencil eraser to push the inserts in,
then check that inserts are aligned correctly (right side up).
Rotary encoder from outside GUI
assembly
Rotary encoder from inside GUI
assembly (note orientation).