Appendix A
A-4
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Block Patient Wye
(Test 1, Step 5)
This step energizes inhalation and exhalation solenoids to autozero the
inhalation and exhalation pressure transducers at ambient pressure. This step
averages eight readings, which must be within 0.8 cmH
2
O of each other.
Diagnostic code
3156
indicates an inhalation pressure transducer (PT3)
autozero failure. Diagnostic code
3157
indicates an exhalation pressure
transducer (PT2) autozero failure.
Red = air flow
Blue = O
2
flow
Black = no flow