Chapter 5
5-2
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Repair
Entering Diagnostic
Mode
To start Esprit ventilator diagnostics, simultaneously press the
ALARM RESET
and
100% O
2
keys on the front panel for approximately five seconds while
turning ventilator power on. At the prompt, touch
OK
to enter diagnostic mode
Figure 5-1: Entering Diagnostic Mode
User Configuration
The user configuration screen (Figure 5-2) allows you to:
•
Change the date and time.
•
Enable or disable the automatic patient circuit compliance
compensation feature.
•
Set the proper altitude.
To view the user configuration screen, touch
User Config
on the diagnostic
screen.
Figure 5-2: User Configuration Screen