Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-43
3. From outside the ventilator, remove the humidifier outlet cover.
Figure 9-49: Humidifier Outlet Cover
4. Remove the two jackscrews from the outside of the receptacle.
Figure 9-50: Removing AC Receptacle Jackscrews
5. Unscrew the other end of the blue neutral wire from the AC receptacle,
and discard the wire.
6. Insert the longer replacement neutral wire (P/N 1012131) into the AC
receptacle: twist the wires if necessary, and check that no copper
wiring extends outside the receptacle after the wire is inserted.
7. Rotate the receptacle and use the jackscrews to secure it into place.
Mains
circuit
breaker
Humidifier
circuit
breaker
Humidifier
outlet
cover
Torque screw
to 8.0 ± 1
in-lbs.
Jackscrews
Humidifier outlet