Performance Verification
Displayed bus voltage is
Backup Battery below
22.3 VDC during
Backup Battery
operation
1. Allow Backup Battery to fully charge (recharge for 8 to 12 hours if
fully depleted).
2. Check the Backup Battery output voltage with a DMM. Replace
battery if the output voltage is below 22.3 VDC.
Cycling ventilator power
results in diagnostic
code or continuous
backup alarm condition
1. Allow Backup Battery to fully charge (recharge for 8 to 12 hours if
fully depleted).
2. Slave-in a known-good Backup Battery.
3. Replace the power supply.
Symptom
Recommended Repair