Chapter 8
Performance Verification
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
8-5
•
Remove the humidifier from the ventilator, if applicable.
•
Visually inspect the exterior of the ventilator for damage. Replace
damaged parts as needed.
•
Remove and inspect the cooling fan filter and air inlet filter. Clean or
replace filter if necessary.
•
View and record the ventilator information:
a.
Simultaneously press the
ALARM RESET
and
100% O
2
keys on
the front panel for approximately five seconds while turning
ventilator power on.
b.
Touch
OK
to enter diagnostic mode (Figure 8-1).
Figure 8-1: Entering Diagnostic Mode
c.
Touch
Software
on the diagnostic screen and record the ventilator
configuration information (Figure 8-2) on the Performance
Verification Data Form (page 8-52).
d.
Touch
User Config
on the diagnostic screen and record the
Altitude (Figure 8-14) on the Performance Verification Data Form
(page 8-52).
e.
Record the elapsed time meter reading on the Performance
Verification Data Form.
f.
Complete a diagnostic report (DRPT) download (Chapter 7).
Touch
Software
to view
ventilator
configuration
Diagnostic
Mode
Touch
User
Config
to view
Altitude