Chapter 5
Repair
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
5-17
7. Disconnect the oxygen source from the ventilator.
8. Return the hardware screen to its initialized state.
Oxygen Regulator
1. Disconnect the bacteria filter, patient circuit tube, and oxygen sensor
tee from the ventilator inspiratory outlet.
2. Connect the oxygen source to the ventilator.
3. Use a pressure gauge capable of measuring 100 PSI, and set the
pressure range to 100 PSI. Zero the gauge.
4. Connect the gauge to the oxygen valve pressure port (Figure 5-12).
Figure 5-12: Oxygen High Pressure Connection
5. Touch
Blower
to de-energize (gray background) the blower.
100 LPM
90 to 110 LPM
120 LPM
108 to 132 LPM
165 LPM
148.5 to 181.5 LPM
0 LPM
0 ± 0.1 LPM
Touch
Oxygen
and Set
the Flow to:
Check that
Analyzer
and
Oxygen
Flow
Displays Read:
Oxygen valve
Connection to
oxygen valve
pressure port