Chapter 5
Repair
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
5-31
High Internal Oxygen
Alarm Test
Follow these steps if a
High Internal O
2
alarm occurs:
1. Disconnect oxygen from the ventilator and wait 5 minutes to allow the
internal oxygen percentage to stabilize.
2. Power up the ventilator in diagnostic mode and touch the
Hardware
button to view the Hardware screen.
3. Confirm that the Internal Oxygen display on the Hardware screen is
between 0.52 to 0.62 V. If the display is outside this range, replace
the sensor PCB
.