Component Removal/Installation
5. Remove the remote alarm jack and washer from the Main PCB/top
enclosure.
Figure 9-59: Remote Alarm Jack (Inside view)
Power Switch
Follow these steps to remove the power switch (Figure 9-60). Reverse removal
steps to install.
NOTE:
Orient the switch as shown when reinstalling.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Disconnect the spade connectors from the power switch terminals.
3. Press the flex clips on the side of the switch and push down to remove
the switch.
Figure 9-60: Power Switch Wire Connections
FIO
2
Connector
Follow these steps to remove the FIO
2
connector (Figure 9-61). Reverse
removal steps to install.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Cut the tie wrap that holds the FIO
2
cable near the sensor PCB.
Red
Brown
Flex clips
Yellow
Orange
Power switch