Chapter 7
7-6
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Ventilator Communications
4. Remove the parallel port adapter from the ventilator back panel.
Figure 7-5: Parallel Port Adapter and Option Enable Button
Setting Up the Serial
Interface for DRPT
Follow these steps to connect the Esprit Ventilator and a PC to create a
diagnostic report (DRPT):
1. Connect a 9-pin male-female null modem RS-232 cable between the
PC and Esprit ventilator. To verify that the cable is a null modem
cable, use a DMM to verify continuity between these DB9 connector
pins at opposite ends of the cable:
2. Power up the ventilator in diagnostic mode: simultaneously press the
Alarm Reset
and
100% O
2
keys for approximately five seconds while
turning ventilator power on.
Option enable
button
Parallel port
adapter
RS-232 Cable Continuity Check
Female DB9 Pin
Male DB9 Pin
1
4
2
3
3
2
4
1 and 6
5
5
6
4
7
8
8
7
9
Not connected