![background image](http://html2.mh-extra.com/html/rands/sgt100a/sgt100a_user-manual_3081093300.webp)
Performing Configuration Tasks
R&S
®
SGT100A
269
User Manual 1176.8674.02 ─ 07
9.5 How to Load and Play Waveform Files
Irrespective of the way a waveform file is generated, you can transfer it to the instru-
ment, load it in the ARB and play it.
To load and play a waveform file from the instrument
1. Transfer an externally created waveform file to the instrument, see
2. Select "Baseband > ARB > General".
3. Set "Waveform Location > Instrument".
4. Select "Load Waveform...". Navigate to the directory the file is stored in, select the
waveform file (e.g.
/var/user/lte_waveform
) and execute "Select".
The "ARB" dialog confirms that the waveform file is loaded.
5. Select "ARB > State > On".
The R&S
SGT processes the waveform file.
To load and play a waveform file from an external computer
1. Connect the R&S
SGT and a Windows PC with an installed R&S SGMA-GUI to the
same network.
2. On the remote PC, select "SGMA-GUI > Instrument Name > Baseband > ARB >
General".
3. Set "Waveform Location > Local file system ".
4. Select "Load Waveform...". Navigate to the directory the file is stored in, select the
waveform file (e.g.
/var/user/lte_waveform
) and execute "Select".
Note:
You can also select an external USB connected to the PC at the available
local file directories.
The "ARB" dialog confirms that the waveform file is loaded.
5. Select "ARB > State > On".
The R&S
SGT processes the waveform file.
9.6 How to Restore the LAN Connection to an Instrument
► If the LAN connection to an instrument configured to use a static IP address is lost,
press the LAN LED on the instrument's front panel for more than 3 seconds to
reset the LAN settings and to set the
How to Restore the LAN Connection to an Instrument