[:SOURce<hw>]:BB:POWer:PEAK?
This command queries the peak power of the baseband signal relative to full scale of
0.5 V (in terms of dB full scale).
Return values:
<Peak>
float
Range:
-145 to 30
Increment: 0.01
*RST:
0
Default unit: dBfs
Example:
BB:POW:PEAK
queries the peak power of the baseband signal.
Usage:
Query only
[:SOURce<hw>]:BB:POWer:RMS?
This command queries the power of the baseband signal relative to full scale of 0.5V
(in terms of dB full scale).
Return values:
<Rms>
float
Range:
-145 to 30
Increment: 0.01
*RST:
0
Default unit: dBfs
Example:
BB:POW:RMS
queries the power of the baseband signal.
Usage:
Query only
11.14.9 SOURce:CORRection Subsystem
The output level is corrected in the
CORRection
subsystem. Correction is performed
by user-defined table values being added to the output level for the respective RF fre-
quency. In the R&S
SGT, this subsystem is used to select, transfer and activate user
correction tables.
Each list is stored as a file. The name of the user correction file can be freely selected.
The file extension
*.uco
is assigned automatically and cannot be changed.
The files can be stored in a freely selectable directory and opened from there. The
default directory is set using command
case of files which are stored in the default directory, only the file name has to be
specified in commands. Otherwise, the complete absolute path has to be specified with
every command. The extension can be omitted in any case.