Parameters:
<LockRequestId>
Number
0
test query to prove whether the instrument is locked
Controller ID
request lock from the controller with the specified Controller ID
Return values:
<Value>
Number
0
request refused; the instrument is already locked to other <Lock
Request Id>, i.e. to another controller
1
request granted
Example:
:LOCK? 12345
Response: 1
:UNL 12345
Usage:
Query only
Manual operation:
See
:UNLock
<UnlockId>
Unlocks an instrument locked to a controller with Controller ID =
<Unlock Id>
.
Setting parameters:
<UnlockId>
Number
Unlock ID which uniquely identifies the controller to the instru-
ment. The value must match the Controller ID
<Lock Request
Id>
.
0
Clear lock regardless of locking state
Usage:
Setting only
Manual operation:
See
11.3 Preset Commands
The preset commands are not bundled into one subsystem. Therefore, they are listed
separately in this section. These specific commands are described in the associated
subsystems.
The following presetting actions are available:
●
Activating the default state of all internal instrument functions (
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. TCP/IP address or reference oscillator source settings.
Preset Commands