Remote Control Commands
R&S
®
SGT100A
335
User Manual 1176.8674.02 ─ 07
:SENSe<ch>[:POWer]:FILTer:TYPE
<Type>
Selects the filter mode. The filter length is the multiplier for the time window and thus
directly affects the measurement time.
Parameters:
<Type>
AUTO | USER | NSRatio
AUTO
Automatically selects the filter length, depending on the mea-
sured value. The higher the power, the shorter the filter length,
and vice versa.
USER
Allows you to set the filter length manually. As the filter-length
takes effect as a multiplier of the measurement time, you can
achieve constant measurement times.
NSRatio
Selects the filter length (averaging factor) according to the crite-
rion that the intrinsic noise of the sensor (2 standard deviations)
does not exceed the specified noise content. You can define the
noise content with command
Note:
To avoid long settling times when the power is low, you
can limit the averaging factor limited with the "timeout" parame-
ter (
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
*RST:
AUTO
Example:
SENS1:FILT:TYPE AUTO
Selects automatic filter selection.
Manual operation:
See
:SENSe<ch>[:POWer]:FREQuency
<Frequency>
Sets the RF frequency of the signal, if signal source "USER" is selected (
).
Parameters:
<Frequency>
float
*RST:
1 GHz
Example:
SENS1:SOUR USER
Selects user-defined source.
SENS1:FREQ 2.44GHz
Sets the RF frequency of the source which is 2.44 GHz.
Manual operation:
See