Signal Generator Settings
R&S
®
SGT100A
224
User Manual 1176.8674.02 ─ 07
When to perform zeroing:
●
During warm up after switching on or connecting the instrument
●
After a substantial change of the ambient temperature
●
After fastening the power sensor module to an RF connector at high temperature
●
After several hours of operation
●
When low-power signals are to be measured, e.g. less than 10
dB above the lower
measurement limit.
●
Switch off the RF power source for zeroing, but do not disconnect it from the power
sensor. This proceeding keeps the thermal equilibrium, and the zeroing process
also compensates the noise that superimposes the measured signal (e.g. from a
broadband amplifier).
Related settings and functions
●
Measurements-related settings, like results, filter, filter length:
●
Sensor-specific information and sensor software update:
●
Software version of the connected power sensor:
●
Acquisition of level correction data:
Chapter 7.7.6.5, "User Correction Settings"
Additional information
See Rohde
&
in section "Power
Meters & Voltmeters" for:
●
R&S
NRP power sensor manual
●
Information on the R&S NRP-Z5 sensor hub and the available accessories.
7.7.8.1
NRP Power Viewer Settings
Access:
► Select "Instrument > RF > Power Viewer".
RF Settings