Remote Control Commands
R&S
®
SGT100A
489
User Manual 1176.8674.02 ─ 07
Return values:
<Result>
0 | 1 | RUNning | STOPped
*RST:
STOPped
Usage:
Query only
:TEST:KEYBoard[:STATe]
<State>
Enable/disable keyboard and LED test state.
Parameters:
<State>
0 | 1 | OFF | ON
:TEST:BBIN?
This command performs a selftest on the baseband input hardware options. Several
analog diagnostics points are checked to verify the correct function of the module.
Return values:
<Bbin>
0 | 1 | OFF | ON
Example:
TEST:BBIN?
Response: "0" on success, "1" on fail
Usage:
Query only
:TEST:BB:CONNection?
Queries the state of the connection between connectors USER1 and USER2.
A
0
=
PASS
response indicates that the connection is established, whereas a response
1
=
FAIL
stands for a faulty connection.
Return values:
<Connection>
0 | 1 | PASS | FAIL
Usage:
Query only
Manual operation:
See
:TEST:EIQMode
<EiqMode>
Triggers a connection test for testing the active external IQ devices.
Parameters:
<EiqMode>
IQIN | IQOut
*RST:
IQOut
TEST Subsystem