Remote Control Commands
R&S
®
SGT100A
405
User Manual 1176.8674.02 ─ 07
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:ODELay
<Delay>
Seds the output delay of the external baseband signal.
Parameters:
<Delay>
float
Range:
0 to 1
Increment: 1E-9
*RST:
0
[:SOURce<hw>]:BBIN:PGAin
<PGain>
This command enters the relative gain for the external baseband signal compared with
the signals of the other baseband sources. The actual gain of the different baseband
signals depends not only on the path gain setting but also on the signal characteristics
such as the crest factor and on the number of used sources. used and on the total RF
output power. The gain affects the signal on the "Baseband" In block output.
Parameters:
<PGain>
float
Range:
-50 to 50
Increment: 1E-3
*RST:
0
Default unit: dB
Example:
BBIN:PGA 3dB
sets the relative gain of 3 dB for the external baseband signal.
[:SOURce<hw>]:BBIN:POWer:PEAK
<Peak>
This command enters the peak level of the external baseband signal relative to full
scale of 0.5 V (in terms of dB full scale).
Parameters:
<Peak>
float
Range:
-60 to 3.02
Increment: 0.01
*RST:
0
Default unit: dBfs
Example:
BBIN:POW:PEAK -2
enters the peak level of -2 dBfs.
Manual operation:
See
SOURce Subsystem