Signal Generator Settings
R&S
®
SGT100A
103
User Manual 1176.8674.02 ─ 07
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:PHASe[:STARt]
Phase Step
Sets the step width that is used to increment the phase.
The resulting phase in the carrier table equals:
Phas n* PhaseStep
–
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:POWer:STEP
Delay Start
Sets the delay of the carrier marked by "Carrier Start".
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay[:STARt]
Delay Step
Sets the step width that is used to increment the delay.
The resulting delay in the carrier table equals:
Dela n* DelayStep
–
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:DELay:STEP
Input Waveform File
Calls the "File" dialog for selecting the inputfile with the I/Q signal to be modulated onto
all carriers of the selected carrier range.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:FILE
Apply Assistant Settings
Transfer the assistant settings to the carrier table.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:MCARrier:EDIT:CARRier:EXECute
Carrier Table
The table displays the settings of all available carriers. Previously applied assistant set-
tings can be further refined. The number of lines corresponds to the number of carriers.
Note:
The phase/deg settings are only valid if optimization of the crest factor is dis-
abled ("Crest Factor Mode" = Off).
Baseband