Remote Control Commands
R&S
®
SGT100A
354
User Manual 1176.8674.02 ─ 07
Example:
BB:ARB:CLOC:SOUR EXT
selects the external clock source. The clock is supplied via the
CLOCK connector.
BB:ARB:CLOC:MODE MSAM
selects clock type Multiple Samples, i.e. the supplied clock has a
rate which is a multiple of the sample rate.
BB:ARB:CLOC:MULT 12
The multiplier for the external clock rate is 12.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:SOURce
<Source>
The command selects the source for the digital modulation clock.
Parameters:
<Source>
INTernal | EXTernal | AINTernal
INTernal
The internal clock reference is used.
EXTernal
The external clock reference is supplied to the connector.
*RST:
INTernal
Example:
BB:ARB:CLOC:SOUR EXT
selects an external clock reference. The clock is supplied via the
connector.
BB:ARB:CLOC:MODE SAMP
enters clock type sample.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:CLOCk:SYNChronization:MODE
<Mode>
Selects the synchronization mode.
This parameter is used to enable generation of very precise synchronous signal of sev-
eral connected R&S SGTs.
Note:
If several instruments are connected, the connecting cables from the master
instrument to the slave one and between each two consecutive slave instruments must
have the same length and type. Avoid unnecessary cable length and branching points.
SOURce Subsystem