Remote Control Commands
R&S
®
SGT100A
398
User Manual 1176.8674.02 ─ 07
Return values:
<Name>
string
Example:
BB:ARB:WSEG:NAME?
queries the currently output segment.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT
<Next>
Selects the segment to be output.
Parameters:
<Next>
float
Range:
0 to 1023
*RST:
0
Example:
MMEM:CDIR '/var/user/waveform'
sets the default directory.
BB:ARB:WSEG:NEXT 2
selects segment 2 to be output.
BB:ARB:TRIG:SMOD NEXT
selects extended trigger mode next, i.e. the segment specified
with command
:BB:ARB:WSEG:NEXT
will be output.
BB:ARB:SEQ AUTO
selects trigger mode Auto.
BB:ARB:WAV:SEL 'multi_wv1'
loads multi segment waveform
multi_wv1.wv
. Generation of
segment 2 starts.
BB:ARB:WSEG:NEXT 3
switched at once to output of segment 3.
Manual operation:
See
[:SOURce<hw>]:BB:ARBitrary:WSEGment:NEXT:EXECute
Triggers manually switchover to the subsequent segment in the mutli segment file. A
manual trigger can be executed only when an internal next segment source
(
BB:ARB:WSEG:NEXT:SOUR INT
) has been selected.
To perform a switchover to any segment within the multi segment file, select the next
segment with the command
BB:ARB:WSEG:NEXT
.
This command is disabled, if a sequencing play list is enabled.
Example:
BB:ARB:WSEG:NEXT:SOUR INT
selects internal next segment source.
BB:ARB:WSEG:NEXT 2
selects segment 2 to be output.
BB:ARB:WSEG:NEXT:EXEC
executes a switchover to the next segment.
SOURce Subsystem