Signal Generator Settings
R&S
®
SGT100A
233
User Manual 1176.8674.02 ─ 07
"Not Signal Valid"
Output of high signal to mark the transition state when frequency and
level change.
"Pulse Video Out"
Available only for USER 2.
Output of the internally generated pulse video (modulating) signal.
The video signal level corresponds to the RF envelope.
"Pulse Gen Ext Trigger"
Available only for USER 2.
Input for an external trigger signal, used to trigger the pulse genera-
tor.
"Pulse Mod Ext Source"
Available only for USER 2.
Input for an externally provided pulse modulation signal.
Used when an external pulse modulator source is provided at the
connector.
Remote command:
Trigger Mode
Selects between continuous pulse modulation or pulse modulation triggered by an
external signal.
"Auto"
Generates the modulation signal continuously.
"Ext Single"
Generates the signal each time an external trigger event occurs.
"Ext Gated"
Generates the signal triggered by an external gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:MODE
Ext. Trigger Input Slope
Available only for "Trigger Mode > Ext Triggered "
Sets the polarity of the active slope of an applied external trigger signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe
Gate Input Polarity
Available only for "Trigger Mode > Ext Gated "
Sets the polarity of the active slope of an applied gate signal.
Remote command:
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity
RF Settings