![background image](http://html2.mh-extra.com/html/rands/sgt100a/sgt100a_user-manual_3081093325.webp)
Network and Remote Control Operation
R&S
®
SGT100A
294
User Manual 1176.8674.02 ─ 07
On the R&S
SGT the LXI functionality is already installed and enabled. Thus, the
instrument can be accessed via any web browser (like the Microsoft Internet Explorer)
to perform the following tasks:
●
Modifying network configurations
●
Remote control of the instrument
●
Performing SCPI remote diagnostics
10.5.1 Default State of the Network Settings
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter
Value
TCP/IP mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP ping
Enabled
Password for LAN configuration
LxiWebIfc
The LAN reset also resets the following parameters for the R&S
SGT:
Parameter
Value
Hostname
<Instrument-specific host name>
Description
Signal generator
Negotiation
Auto detect
VXI-11 discovery
Enabled
The LAN settings are configured using the instrument's
.
10.5.2 LXI Browser Settings
To access the instrument via the web browser:
► Type in the instrument's host name or IP address in the address field of the
browser on your PC, for example "http://10.111.0.97".
Note:
Do not add the missing zeros in the IP address, while opening the instru-
ment home page.
LXI Configuration