Instruction Manual

                    

Model 2099-1318 

10MHz Source/Inserter, Redundant Power

 

August 2017, Rev. A 

ALARM

LNB1

DC ON

REF ON

LNB2

MENU

EXECUTE

ALARM

PSA
PSB

EXT REF

REMOTE
OVEN

MODEL 2099

SOURCE / INSERTER

LNB1  +13.2V @  0.249A

LNB2  +18.1V @  0.251A

C

ROSS

 T

ECHNOLOGIES 

INC

Data, drawings, and other material contained herein are proprietary to Cross Technologies, Inc., 

but may be reproduced or duplicated without the prior permission of Cross Technologies, Inc. 

for purposes of operating the equipment. Printed in USA.

When ordering parts from Cross Technologies, Inc., be sure to include the equipment 
model number, equipment serial number, and a description of the part.

C

ROSS 

T

ECHNOLOGIES, INC.

6170 Shiloh Road

Alpharetta, Georgia 30005

(770) 886-8005

FAX (770) 886-7964

Toll Free 888-900-5588

WEB:  www.crosstechnologies.com 

E-MAIL:  [email protected]

Summary of Contents for 2099-1318

Page 1: ...oprietary to Cross Technologies Inc but may be reproduced or duplicated without the prior permission of Cross Technologies Inc for purposes of operating the equipment Printed in USA When ordering parts from Cross Technologies Inc be sure to include the equipment model number equipment serial number and a description of the part CROSS TECHNOLOGIES INC 6170 Shiloh Road Alpharetta Georgia 30005 770 8...

Page 2: ...e year after shipment to customer Cross Technologies Inc s obligation under this warranty is limited to repairing or at Cross Technologies Inc s option replacing parts subassemblies or entire assemblies Cross Technologies Inc shall not be liable for any special indirect or consequential damages This warranty does not cover parts or equipment which have been subject to misuse negligence or accident...

Page 3: ...ontrol allows remote configuration of front panel commands and monitoring LNB1 and LNB2 voltage and current Parameter selection and each LNB voltage and current appear on the LCD display Connectors are BNC female for RF and 10 MHz input and output signals Redundant AC power is 100 240 10 VAC 47 63 Hz The chassis is 1 RU 12 deep ALARM LNB1 DC ON REF ON LNB2 MENU EXECUTE ALARM PSA PSB EXT REF REMOTE...

Page 4: ...Switch or M C Select External Yellow LED Internal External 10 MHz Select Control Indicators 149 149 140 110 10 kHz 1 kHz 100 Hz 10 Hz dBC Hz Phase Noise Frequency TABLE 1 0 2099 1318 Source Inserter 0 5 ppm Tuning Adjust 0 01 ppm Warm Up 1 Hour 0 1 ppm Warm Up 4 Minutes 0 1 ppm Aging per year 0 001 ppm Aging per day 0 01 ppm max over temp Stability Oscillator 30 dBC 40 dBC Typ Harmonics Output 30 ...

Page 5: ...ubject to change without notice Cross Technologies Inc 2017 50Ω N Type NN 75Ω F Type F 75Ω BNC Ethernet with TCP IP Telnet Ethernet with Web Browser and SNMP Ethernet with Web Browser Connectors Impedance Common Interface Standard RS232 Options B W28 W18 W8 External 10 MHz Reference Auto Detect and Restore E ...

Page 6: ... 3 4 5 6 7 8 9 2 1 3 4 5 6 7 8 9 M C Cable Diagram Cross Technologies Frequency Converters and Frequency Sources Connector Rear panel DB 9 female TABLE 2 1 J19 TABLE 2 1 J19 Pinouts DB9 Pin Function 1 Rx RS 485 2 Rx RS 232C RS485 3 Tx RS 232C RS485 4 Tx RS485 5 GND 6 Alarm Relay Common 7 Alarm Relay Open ALARM 8 Not Used 9 Alarm Relay Closed ALARM 2099 1318 Manual Rev A Page 6 08 23 17 ...

Page 7: ...C voltage is inserted on the LNB2 loop through x 0 if DC voltage is disabled on the LNB2 loop through Get LNB2 10 MHz Reference Insert State aaSB returns aaSBx where x 1 if the 10 MHz reference is inserted on the LNB2 loop through x 0 if the 10 MHz reference is disabled on the LNB2 loop through Get LNB1 Inserted Voltage and Current aaSJ returns aaSJxxxxx yyyyy where xxxxx Measured DC voltage inser...

Page 8: ...1 I status 0 ok 1 current too high h LNB2 V status 0 ok 1 voltage too high 2 voltage too low i LNB2 I status 0 ok 1 current too high j Internal Reference Mode status 0 ok 1 internal reference low and Internal Reference Mode is selected k External Reference Mode status 0 ok 1 external reference low and External Pass Reference Mode is selected l External Pass Auto Reference Mode status 0 ok 1 extern...

Page 9: ...oltage to 18 VDC note When the LNB1 voltage changes the LNB1 DC insert will automatically turn off to safeguard any external equipment The LNB1 DC insert will have to be set to on after any LNB1 voltage change Set LNB2 DC Voltage Insert aaCLx where x 1 to enable DC Voltage on the LNB2 loop through x 0 to disable DC Voltage on the LNB2 loop through Set LNB2 10 MHz Reference Insert aaCBx where x 1 t...

Page 10: ...undant switching 18 VDC power supplies provides power for the assemblies The 2099 1318 10MHz Source Inserter can be secured to a rack using the 4 holes on the front panel Figure 2 0 shows how the 2099 1318 10MHz Source Inserter is assembled AC A AC B FIGURE 2 0 2099 1318 10MHz Source Inserter Mechanical Assembly 2099 1318 Manual Rev A Page 10 08 23 17 ...

Page 11: ...2150 MHz Capacitively coupled and looped from LNB2 RF input 50Ω BNC female connector F1 LNB2 Fuse 5A Fast Blo 1 4 Fuse J6 10MHz REF OUTPUT 10MHz reference output 3 dBm 3 dB 50Ω 75Ω BNC female connector J8 LNB1 RF OUT 950 2150 MHz Capacitively coupled and looped from LNB1 RF input 50Ω BNC female connector DS2 LNB1 ALARM DC Lights yellow when LNB1 voltage is present on the RF OUT J8 center pin J7 LN...

Page 12: ...ndicates 10 MHz insertion on LNB1 RF output DS2 LNB2 Alarm Indicates voltage insertion out of range or fuse alarm DC ON Indicates DC presence on LNB2 RF input REF ON Indicates 10 MHz insertion on LNB2 RF input DS3 Summary Alarm Indicates a problem with an internal subsystem or a problem with the external reference input PSA AC power on power supply ʻAʼ PSB AC power on power supply ʻBʼ DS4 External...

Page 13: ... or Ext Lock Auto in which to operate the unit 7 AC FUSE The fuse is a 5 mm X 20 mm 2 amp slow blow Type T and is inserted in the far slot in the drawer below the AC input as shown in Figure 2 3 There is a spare fuse in the near slot If a fuse continues to open the power supply is most likely defective 8 AC FUSE The fuse is a 5 mm X 20 mm 2 amp slow blow Type T and is inserted in the far slot in t...

Page 14: ...Menu 8 Adjust Internal Reference Frequency Offset Menu 9 Select Remote Operation On Off Menu 10 Select Remote Interface 232 422 or 485 Menu 11 Select RS485 Address Menu 12 Measured 18 VDC From Power Supplies AC A AC B Menu 13 Measured 13 VDC From DC to DC Converters Save Menu When R is selected from any above menu or at the end Alarm indications appear on the LED see Figure 2 2 All program changes...

Page 15: ...s now operational and ready for any changes the operator may desire 2 5 3 Control Switches 1 Menu Execute Any change to the programming of the unit must be initiated by pressing the Menu Execute switch and completed by pressing the Menu Execute switch 2 Horizontal Switch This switch is mounted so its movement is horizontal and moves the cursor left or right 3 Vertical Switch This switch is mounted...

Page 16: ... Supplies A B Menu 13 Measured 13 VDC From DC to DC Converters A B Save When R is selected from any above menu or at the end LNB1 REFERENCE INSERT ON SCROLL PUSH BUTTON R R LNB2 DC VOLTS 18 SCROLL SCROLL PUSH BUTTON R SCROLL SCROLL PUSH BUTTON R SCROLL SCROLL PUSH BUTTON R SCROLL SCROLL PUSH BUTTON REF MODE INTERNAL R LNB1 13 VDC INSERT OFF SCROLL SCROLL PUSH BUTTON R ON POWER UP 1 IP Addr 192 168...

Page 17: ...gh LNB2 v insert low LNB2 v Insert ok LNB2 v Insert off STATUS 7 LNB1 i insert high LNB1 i insert low LNB1 i Insert ok LNB1 i Insert off STATUS 6 LNB1 v insert high LNB1 v insert low LNB1 V Insert ok LNB1 V Insert off STATUS 5 DC conv b high DC conv b low DC Conv B ok STATUS 4 DC conv a high DC conv a low DC Conv A ok STATUS 3 PSB 18V high PSB 18V low PSB 18V ok STATUS 2 PSA 18V high PSA 18V low P...

Page 18: ...be detected before this voltage is monitored for alarm purposes 5 LNB1 V Monitor This is the DC voltage that is inserted at the LNB1 RF loop through The LNB1 alarm indicator and alarm contact closure will be set if the selected voltage either 13V or 18V is too high or too low If 13V is selected then the unit will alarm if the voltage is greater than 0 14V or less than 0 13V If 18V is selected then...

Page 19: ... less than 5 dBm 1 dBm The external reference indicator will always be on in this mode 4 External Lock Reference Mode The summary alarm indicator and contact closure will be set if the external reference power is less than 5dBm 1 dBm The summary alarm indicator and contact closure will be set if the internal reference cannot lock to the external reference The external reference indicator will alwa...

Page 20: ...equipment is not compromised Additional space between units may be required E Circuit Overloading Consideration should be given to the connection of the equipment to the supply circuit and the effect that overloading of circuits could have on over current protection and supply wiring Appropriate consideration of equipment name plate rating should be used when addressing this concern F Reliable Ear...

Page 21: ...S INC 6170 Shiloh Road Alpharetta Georgia 30005 770 886 8005 FAX 770 886 7964 Toll Free 888 900 5588 WEB www crosstechnologies com E MAIL info crosstechnologies com Printed in USA 2099 1318 Manual Rev A Page 21 08 23 17 ...