Remote Control Commands
R&S
®
SGT100A
404
User Manual 1176.8674.02 ─ 07
Example:
BBIN:OFFS:Q -0.1PCT
sets -0.1 %FS DC offset to the Q component of the external
input signal.
[:SOURce<hw>]:BBIN:OLOad:HOLD:RESet
This command executes a reset of the Overload Hold State and LED.
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:RES
resets the overload hold state.
Usage:
Event
[:SOURce<hw>]:BBIN:OLOad:HOLD:STATe?
This command queries an overload since the last reset for evaluating the measure-
ment. The Overload Hold state can be reset via the "Reset" command, or is reset auto-
matically by starting a new measurement ("Auto Level Set") or by setting new values
(Crest Factor, Peak Level, Level).
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:HOLD:STAT?
queries the overload state.
Response: 0
the AC/DC converter has not been overloaded.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:OLOad:STATe?
This command queries the current overload state of the A/D converter.
Return values:
<State>
0 | 1 | OFF | ON
*RST:
0
Example:
BBIN:MODE DIG
selects the digital baseband input mode.
BBIN:OLO:STAT?
queries the overload state.
Response: 0
The AC/DC converter is not overloaded.
SOURce Subsystem