Remote Control Commands
R&S
®
SGT100A
334
User Manual 1176.8674.02 ─ 07
Parameters:
<NSRatio>
float
Range:
0.001 to 1
Increment: 0.001
*RST:
0.01
Example:
SENS1:FILT:TYPE NSR
Selects fixed noise filter mode.
SENS1:FILT:NSR 0.2
Sets a noise content of 0.2.
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe
<MTime>
Sets an upper limit for the settling time of the auto-averaging filter in the
NSRatio
mode and thus limits the length of the filter. The filter type is set with command
.
Parameters:
<MTime>
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
Example:
SENS1:FILT:TYPE NSR
Selects fixed noise filter mode.
SENS1:FILT:NSR .2
Sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
Limits the settling time to 5 seconds.
Manual operation:
See
:SENSe<ch>[:POWer]:FILTer:SONCe
Starts searching the optimum filter length for the current measurement conditions. You
can check the result with command
:SENS1:POW:FILT:LENG:USER?
in filter mode
USER
:SENSe<ch>[:POWer]:FILTer:TYPE
).
Example:
SENS1:FILT:TYPE USER
Selects user filter mode.
SENS1:FILT:SONC
Activates the search for the optimum filter length.
SENS1:FILT:LENG?
Returns the found optimum filter length.
Response: 128
Usage:
Event
Manual operation:
See
SENSe, READ, INITiate and SLISt Subsystems