Tablet
Continued...
6.5 - 8.4
(pH)
Name on device: 38-pH-MR-tab
(38)
pH-value (MR)
Note(s)
-> Use of this test procedure and reagent on water samples with pH value
outside of the 6.5 - 8.4 range can lead to incorrect test results. If you are not
sure we recommend a control measurement using e.g. an electronic meter
(pH 0-14).
-> If the alkalinity (or carbonate hardness) of your water sample is lower than
KS4.3 = 0.07 mmol/l (= 35 mg/l CaCO3) this can lead to incorrect test
results.
-> Depending on the salt content of your sample, the measurement result must
be manually corrected according to the following scheme:
1 molar = -0.21 pH
2 molar = -0.26 pH
3 molar = -0.29 pH
with: 1 mol of salt (NaCl) = 5.8% = 58.4 g/l
Содержание 1.0 MULTITEST
Страница 1: ......
Страница 21: ...Menu MEM Memory left open for technical reasons MEM 5 MEM 5 ...
Страница 23: ...INFO 2 INFO 2 Menu INFO left open for technical reasons ...
Страница 33: ...left open for technical reasons ...
Страница 276: ...243 243 Tablet Note s Conversion to Na SO 2 3 Continued Name on device 53 Sulphite LR tab 53 Sulphite LR 0 5 mg l SO 3 ...
Страница 323: ...PrimeLab Desktop Assistant PDA 9 PDA 9 left open for technical reasons ...
Страница 333: ...PrimeLab Desktop Assistant PDA 19 PDA 19 left open for technical reasons ...
Страница 340: ...PrimeLab Desktop Assistant Menu Account data Print test results report PDA 26 PDA 26 ...
Страница 342: ...PrimeLab Desktop Assistant Menu Account data Generate dosage recommendations PDA 28 PDA 28 ...
Страница 343: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 29 PDA 29 ...
Страница 347: ...Page empty for technical reasons PrimeLab Desktop Assistant PDA 33 PDA 33 ...